RWCRealWorldCalculatorTurn real numbers into real insights.

Measurement Uncertainty Calculator

Combine repeated readings with resolution and calibration uncertainty.

The interactive calculator loads above this explanation when JavaScript is available.

How to use the measurement uncertainty calculator

Combine repeated readings with resolution and calibration uncertainty. Enter the known values, choose units where offered, and read the live result and breakdown.

Worked example

This example uses the following inputs. Currency amounts are illustrated in USD.

Repeated readings in the same unit
10.01, 10.02, 10.00, 10.01, 10.02
Instrument resolution increment
0.01
Calibration expanded uncertainty
0.02
Calibration coverage factor
2
Output coverage factor
2

Expanded uncertainty: 0.0221 same unit

Method

Repeatability u = s/√n. Resolution assumes rounding within ±increment/2 with a rectangular distribution: u = increment/√12. Calibration u = quoted expanded uncertainty / its coverage factor. Combine independent components by root sum of squares.

What to watch for

Same quantity and unit, independent components, corrected known bias. No automatic confidence-percentage claim. Avoid double-counting resolution if already included in another component. This does not test conformance to manufacturing tolerances.

Assumptions and references

Same quantity and unit, independent components, corrected known bias. No automatic confidence-percentage claim. Avoid double-counting resolution if already included in another component. This does not test conformance to manufacturing tolerances.

NIST: combining uncertainty ↗

Formula implementation: v0.8.0. Examples are checked against the calculation engine; specialist professional review is not claimed. How we check calculations.

Report a problem with this calculator

Related calculators